Why is HP 5370/5345 Time Interval Measuring less accurate tr
Electronics Forum Index Electronics
Circuits, theory, electrons and discussions.
 
 FAQFAQ   MemberlistMemberlist     RegisterRegister 
 ProfileProfile   Log in to check your private messagesLog in to check your private messages   Log inLog in 
 
Google
 
Web ElectronicsHelp.net
Why is HP 5370/5345 Time Interval Measuring less accurate tr

 
Post new topic   Reply to topic    Electronics Forum Index -> Equipment
Author Message
Peter, PE1ECM
Guest





Posted: Fri Jan 28, 2005 2:07 am    Post subject: Why is HP 5370/5345 Time Interval Measuring less accurate tr Reply with quote

I use a HP 5345B and a HP 5370A for time interval measuring ( separate
as well as common mode ).
The 5345A has a jitter stability ( useable resolution ) of 2 nS and
the 5370A a jitter stability of 20 pS.
Both the instruments behave OK.
However, I see that the stability of the period is far better than the
accuracy of the time interval.

Whatever tweaking I have done and whatever signal source I use ( e.g.
a e-11 frequency reference )and at whatever frequency I perform the
measurements, any time interval will be no more stable than about 5e5.
Thus, measuring ( 5370A )( common TI mode ) a 10 Mhz frequency
reference signal the stability shows about +/- 40 pS jitter which is
very good and far within the specs.

Measuring at 1 Mhz ( reference locked ) however, the stability is +/-
some hundreds of pS.
And measuring at 10 Khz , the stability is +/- 50 nS

I get about the same ( relative ) results when using the 5345B instead
of the 5370A ( be it that the jitter stability and resolution of the
5345B is 100 x less ).

However, measuring the period ( instead of the TI mode ) I get the
expected results of about e-11( which is cf the specs if locked to a
better than e-11 freq.ref ).

In the frequency modes both the instruments perfom far within specs (
stability resolution better than e-11 ).

What is it that TI measurement ( in these particular HP gear )is so
much less stable than period measurent ?

Since both the instruments show this very phenomenum and for both even
in the same relationship at varying frequencies, I guess this must be
system feature and no flaw.

Any HP guru in this group ?

I bent my mind for days but I cannot resolve this one.
Thanks for help.

Peter, PE1E ex PE1ECM

Back to top
Jim Adney
Guest





Posted: Fri Jan 28, 2005 12:07 pm    Post subject: Re: Why is HP 5370/5345 Time Interval Measuring less accurat Reply with quote

On 27 Jan 2005 12:07:12 -0800 pe1ecm@pmvandaalen.myadsl.nl (Peter,
PE1ECM) wrote:

Quote:
What is it that TI measurement ( in these particular HP gear )is so
much less stable than period measurent ?

There is a built-in time uncertainty of one internal clock oscillator
period in any counter. This is the resolution of your counter.

For an TI measurement you are measuring a single event, so your
resolution on that measurement is your uncertainty.

For a period measurement, many instruments will measure over N periods
and then divide by N to get the period. The same resolution above
applies to the total time measured, but this makes the uncertainly for
a single period 1/N times the measured uncertainty.

Often you can select N when you set up the instrument; it will be
related to the gate time that you use when you measure frequency. Just
as your frequency measurement gets better if you average over a longer
gate period, so will your period measurement.

-
-----------------------------------------------
Jim Adney jadney@vwtype3.org
Madison, WI 53711 USA
-----------------------------------------------
Back to top
 
Post new topic   Reply to topic    Electronics Forum Index -> Equipment All times are GMT
Page 1 of 1

 
You cannot post new topics in this forum
You cannot reply to topics in this forum
You cannot edit your posts in this forum
You cannot delete your posts in this forum
You cannot vote in polls in this forum



Home & Living New Topics
Contact Us
Powered by phpBB